Sticking for the perception of Creating goods of high quality and making good friends with people today from all around the world, we constantly set the interest of shoppers to begin with for Portable Digital X Ray, Filter In X Ray Machine, Xray Voltage, We Sincerely welcome customers from all over the world to visit us, with our multifaceted cooperation and work together to develop new markets, create win-win brilliant future.
OEM High Quality X Ray Leaded Glass Manufacturer - Dental X-ray Tube CEI Ox_70-m – Sailray Detail:

KL1-0.8-70 Stationary anode X-Ray Tube is especially designed for intra-oral dental x-ray unit and available for a nominal tube voltage with self-rectified circuit.
KL1-0.8-70 tube has one focus.
The integrated high quality tube with glass design has one super imposed focal spot and a reinforced anode.
The high anode heat storage capacity ensures a wide range of applications for intra-oral dental application. A special designed anode enables an elevated heat dissipation rate which leads to a higher patient throughput and a longer product life. A constant high dose yield during the entire tube life is ensured by the high density tungsten target. Ease of integration into system products is facilitated by extensive technical support.

KL1-0.8-70 Stationary anode X-Ray Tube is especially designed for intra-oral dental x-ray unit and available for a nominal tube voltage with self-rectified circuit.

Nominal Tube Voltage 70kV
Nominal Inverse Voltage 85kV
Nominal Focal Spot 0.8 (IEC60336/1993)
Max. Anode Heat content 7000J
Max. Current Continuous Service 2mA x 70kV
Max. Anode Cooling Rate 140W
Target Angle 19°
Filament Characteristics 1.8 – 2.2A, 2.4 – 3.3V
Permanent Filtration Min. 0.6mm Al / 50 kV(IEC60522/1999)
Target Material Tungsten
Nominal Anode Input Power 840W

details

Elevated anode heat storage capacity and cooling
Constant high dose yield
Excellent lifetime

Prior to usage, season the tube in accordance with the seasoning schedule given below until the required tube voltage is attained. Example given – needs to revise by the manufacturer and specified in the data sheet of the part:

Initial incoming seasoning and seasoning schedule for idle period (more than6months)Circuit:

When the tube current is unstable in seasoning, immediately switch off the tube voltage and after an interval of 5 minutes or more, increase the tube voltage gradually from the low voltage while making sure that the tube current is stable. The withstand voltage performance of the tube unit will be lowered as the exposure time and the number of operation increase. Stain-like impact traces might appear on the x-ray tube target surface by slight discharge in the course of the seasoning. These phenomena are one process to recover the withstanding voltage performance at that time. Therefore, if it is in stable operation at the maximum tube voltage of seasoning subsequent to them, the tube unit can be used without any interference to its electrical performance which in use.

Cautions

Read the cautions before using the tube

 

X-ray tube will emit X –ray when it is energized with high voltage, Special knowledge should be required and cautions need to be taken when handling it。

 

1. Only a qualified specialist with X-Ray tube knowledge should assemble, maintain and remove the tube。

2. Sufficient care should be taken to avoid strong impact and vibration to the tube because  it is made of fragile glass。

3. Radiation protection of the tube unit must be sufficiently taken。

4. The minimum sorce-skin distance (SSD) and the minimum filtration should match the regulation and meet the standard。

5. The system should have proper overload protection circuit, the tube may be damaged due to only one overload operation。

6. When any abnormalities are found during operation, immediately switch off the power supply and contact the service engineer。

7. if the tube is with a lead shield, to dispose of lead shield must meet government regulations。


Product detail pictures:

OEM High Quality X Ray Leaded Glass Manufacturer - Dental X-ray Tube CEI Ox_70-m – Sailray detail pictures

OEM High Quality X Ray Leaded Glass Manufacturer - Dental X-ray Tube CEI Ox_70-m – Sailray detail pictures

OEM High Quality X Ray Leaded Glass Manufacturer - Dental X-ray Tube CEI Ox_70-m – Sailray detail pictures


Related Product Guide:

Our pros are lessen prices,dynamic sales team,specialised QC,sturdy factories,top quality services and products for OEM High Quality X Ray Leaded Glass Manufacturer - Dental X-ray Tube CEI Ox_70-m – Sailray , The product will supply to all over the world, such as: Bogota, Stuttgart, Singapore, We firmly think that we have the full capability to present you contented merchandise. Wish to collect concerns within you and build a new long-term synergy romantic relationship. We all significantly promise:Csame excellent, better selling price; exact selling price, better quality.

Minimum Order Quantity: 1pc

Price: Negotiation

Packaging Details: 100pcs per carton or customized according to the quantity

Delivery Time: 1~2 weeks according to the quantity

Payment Terms: 100% T/T in advance or WESTERN UNION

Supply Ability: 1000pcs/ month

  • This is the first business after our company establish, products and services are very satisfying, we have a good start, we hope to cooperate continuous in the future!
    5 Stars By Novia from Malaysia - 2018.06.21 17:11
    This manufacturers not only respected our choice and requirements, but also gave us a lot of good suggestions, ultimately, we successfully completed the procurement tasks.
    5 Stars By Rita from Mecca - 2018.12.14 15:26
    Write your message here and send it to us